LX-077TM Standard (Manual Control
Type)
|
Items
|
Technical Specifications
|
Base
|
Granite overall base, dimensions: 620mmX360mmX110mm
|
Eyepiece
|
10X wide-angle eyepiece with large field of view, the field of view number is Ф22mm
|
Objective
|
Type
|
Magnification
|
Numerical Aperture
|
Working Distance(mm)
|
Infinity
bright&dark field of view Plan Achromatic Objectives
|
5X
|
0.12
|
8.4
|
10X
|
0.25
|
9.3
|
20X
|
0.4
|
7.2
|
50X
|
0.7
|
2.5
|
Viewing Head
|
Erect image observation, binocular interpupillary distance
adjustment range: 53~75mm, 100% push-pull photography.
|
Focusing Arm
|
Manual coarse and fine coaxial focusing, the minimum fine movement
grid value: 0.002mm, the coarse movement is adjustable, the effective
focusing range is 0 ~ 8mm, and the verticality in the axial plane is not more
than 0.05mm.
|
Nosepiece
|
Five-hole light and dark field converter (thread specification:
M26X0.705), with 1 objective lens adapter (conversion thread specification:
4/5″X1/36″)
|
Working Platform
|
Material: High-quality gray cast iron, treated with aging stress
relief.
|
Maximum effective moving range:
200mmX200mm
|
The verticality of vertical and horizontal
movement is not more than 0.05mm/200mm (measured without load).
|
The parallelism between the glass work
surface and the vertical and horizontal moving plane is not more than 0.03mm
within the effective observation range (measured without load).
|
The maximum carrying weight is not more
than 10 kg.
|
Data Processor
|
DP-100 data processor, can be connected with computer in parallel,
synchronous raster signal output
|
The indication error is not more than
0.003mm, and the return error is not more than 0.005mm.
|
Working voltage: 220V 50/60HZ
|
Illumination
|
Falling illumination adopts 12V50W halogen lamp, brightness and
center position can be adjusted.
|
Falling illuminator built-in aperture
diaphragm, field diaphragm, color filter conversion device, bright and dark
field of view switching device, polarized light observation device.
|
Transmission lighting adopts high
brightness white LED lamp with adjustable brightness.
|
Camera Device
|
3 million pixel high-definition color CMOS chip and LCD display,
with manual/automatic photo, data storage and output functions.
|
|
Optional Accessories
|
Eyepiece
|
10X large field of view, reticle eyepiece, reticle value 0.1mm,
field of view Ф20mm
|
Objective
|
Type
|
Magnification
|
Numerical Aperture
|
Working Distance(mm)
|
Infinity
bright&dark field of view Plan Achromatic Objectives
|
40X
|
0.6
|
3
|
60X
|
0.75
|
1.9
|
80X
|
0.8
|
0.8
|
100X
|
0.85
|
0.4
|
Viewing Head
|
Three-position trinocular head, 100% eyepiece observation,
simultaneous visual and photographic spectroscopic observation (spectroscopy
ratio of 5:5 or 2:8), 100% photographic observation, can be freely converted
by switching the lever.
|
DIC Device
|
Differential interference phase contrast puller assembly for 5X,
10X, 20X bright and dark field objectives.
|
Photo Device
|
DV-130/300/500/1000 color digital microscope camera, with general
microscope image measurement and analysis software.
|
0.5X, 1X, 2.5X, 2.5X~4X camera connector
|